Rice Lake Ranger 5 Load Cell Simulator
Standard Features
- ABS plastic case, weighs less than 1 lb
- Fixed rotary switch, -0.5 mV/V to 4.5 mV/V in nine steps of 0.5 mV/V per step
- 10-turn vernier knob with locking graduated dial
- Vernier selection of four ranges: Off, Fine, Medium and Coarse
- Ideal tool for calibrating indicator A/D offset and gain setting
- A/D calibration steps for -0.5 mV/V for offset and 4.5 mV/V for gain
- Used in testing, troubleshooting and calibrating digital weight indicator
Designed for Advanced Troubleshooting
The Ranger 5 variable range simulator is a full-functioning simulator, well suited for more advanced troubleshooting and calibration needs.
Load cell simulator from -0.5 mV/V to 4.5 mV/V and resolution of 0.5 mV/V. Ideal for testing and diagnostics of digital weight indicators, calibration of A/D converters.
Technical Specifications
|
Fixed Steps: Vernier: Impedance: |
Nonlinearity: Connectors: Temperature Coefficient: |
Dimensions (L × W × H): |
Calibration without Weights
Theoretical calibration of weighing systems without the use of sample masses! Useful in high-capacity weighing systems such as silos and hoppers.
A/D converter calibration:
Adjust the A/D converter and enter the mV/V value of the fully loaded cell signal indicated on the calibration certificate on the weight indicator. This load cell simulator is manufactured with high-precision electronic components.
Parts
| Part # | Description |
|---|---|
| Available Versions | |
| 68031 | Simulator,Ranger 5 Selectable .5 MV Steps w/ Vernier |
Options
| Part # | Description |
|---|---|
| Cables | |
| 21116 | Cable,Interface Simulator I 3 |